Model Number 9-ASD-022
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
On (b)(6) 2020, a 22mm amplatzer septal occluder (lot: 7293362) was selected for implant.Once inside the patient, it was observed when opening the left disc of the 9-asd-022, the device took the cobra shape.The device was not released from the cable and was exchanged with a 22 mm amplatzer septal occluder (lot: 7264061).The device was successfully implanted.The patient remained hemodynamically stable throughout the procedure.
 
Manufacturer Narrative
The reported event of the device deploying in a deformed "cobra" shape could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10885425
MDR Text Key217703423
Report Number2135147-2020-00499
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010182
UDI-Public00811806010182
Combination Product (y/n)N
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-ASD-022
Device Catalogue Number9-ASD-022
Device Lot Number7293362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received11/23/2020
Supplement Dates Manufacturer Received12/22/2020
Supplement Dates FDA Received12/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-