Model Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver exhibited an emergency battery error alarm.
|
|
Manufacturer Narrative
|
The emergency battery smbus data was reviewed and showed no permanent faults or diminished capacity that would indicate a malfunction of the emergency battery.The companion 2 driver was plugged in to a/c power for an extended period of time; over 48 hours.This extra charging time allowed the emergency battery to fully recover and subsequently the driver functioned as intended and the driver passed all steps in the companion 2 driver final assembly functional test procedure.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|