Model Number 397002-001 |
Device Problems
Leak/Splash (1354); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(4); initial.
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver exhibited a left pressure incorrect alarm.
|
|
Manufacturer Narrative
|
Visual inspection revealed a missing o-ring on the male cpc connector port used to connect the left driveline to the main valve assembly.During investigation testing, the reported issue was confirmed and replicated.The missing o-ring prevented the driver from attaining/maintaining set pressure requirements during testing.The root cause of the reported left pressure incorrect alarm was determined to be a missing o-ring on the cpc connector.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.Ce 5412 (b)(4) follow-up report 1.
|
|