Device evaluation confirmed the reported issue as the objective cover glass was found to be broken.The optical fiber system were found with chipped lens, dirt in objective unit was observed.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Based on device evaluation results, physical damages to the device probable cause could be attributed to user mishandling.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Olympus will continue to monitor complaints for this device.
|