Model Number M3-30A
Device Problems Break (1069); Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation confirmed the reported issue as the objective cover glass was found to be broken.The optical fiber system were found with chipped lens, dirt in objective unit was observed.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Based on device evaluation results, physical damages to the device probable cause could be attributed to user mishandling.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects.Olympus will continue to monitor complaints for this device.
 
Event Description
It was reported that during reprocessing the device was found with damaged distal tip.There was no patient involvement on this event.No user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Type of Device
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
93 north pleasant st.
norwalk, OH 44857
9013785969
MDR Report Key10886592
MDR Text Key222023680
Report Number1519132-2020-00100
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K890328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3-30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-