U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER CRIBRIFORM OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER CRIBRIFORM OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-MF-030
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 a 30mm amplatzer cribiform device was selected for a procedure.During deployment the left disc was not deployed well.The procedure was completed by using a different device.The patient remained stable through out and post procedure.There was no clinically significant delay in the procedure.
 
Manufacturer Narrative
The reported event of a deformed deployment was confirmed.Following the simulated deployment, the bulbous shape returned to normal and met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.A capa was initiated for further investigation and did not identify a product quality issue.However, corrective actions to further enhance performance are being pursued per internal operating procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER CRIBRIFORM OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10886734
MDR Text Key220968899
Report Number2135147-2020-00471
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010304
UDI-Public00811806010304
Combination Product (y/n)N
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-ASD-MF-030
Device Catalogue Number9-ASD-MF-030
Device Lot Number7371527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Was the Report Sent to FDA? No
Date Manufacturer Received12/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-