Catalog Number A1059 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.Linked to mfg report number (b)(4).
|
|
Event Description
|
This is 2 of 2 reports.A faciliy reported the mayfield skull clamp was hard to lock.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
|
|
Manufacturer Narrative
|
Unique device identifier (udi): (b)(4).The reported complaint was confirmed via inspection of the returned a1059 mayfield skull clamp.Unit received requiring replacement of the index knob and replacement of worn internal parts.This unit is beyond integra's 7 years recommended life cycle (manufactured in 2008).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|