• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; N/A Back to Search Results
Catalog Number A1059
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.Linked to mfg report number (b)(4).
 
Event Description
This is 2 of 2 reports.A faciliy reported the mayfield skull clamp was hard to lock.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
 
Manufacturer Narrative
Unique device identifier (udi): (b)(4).The reported complaint was confirmed via inspection of the returned a1059 mayfield skull clamp.Unit received requiring replacement of the index knob and replacement of worn internal parts.This unit is beyond integra's 7 years recommended life cycle (manufactured in 2008).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10891809
MDR Text Key218007516
Report Number3004608878-2020-00694
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number084
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/24/2020
Supplement Dates Manufacturer Received12/16/2020
Supplement Dates FDA Received01/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-