Catalog Number A1059 |
Device Problem
Device Slipped (1584)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported the locking mechanism of the mayfield skull clamp was not engaging, or slipping.The issue was noted during an internal service.There was no patient involvement, or surgery delay.
|
|
Manufacturer Narrative
|
Udi: (b)(4).Device history record (dhr) - the dhr shows no abnormalities related to the reported failure.The reported complaint was confirmed from the visual evaluation of the returned mayfield skull clamp.The lock had movement and a residue buildup is present , the index knob and the lock had worn parts.Replacements, repair and pm is required at this time.The observed condition is likely caused by wear and tear / improperly handling of the device.The definite root cause cannot be reliably determined.
|
|