Catalog Number A1059 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.Linked to mfg report number: 3004608878-2020-00703.
|
|
Event Description
|
This is 2 of 2 reports.A facility reported the locking knob of the mayfield skull clamp was very loose, and unlocks easily.It is unknown if there was patient involvement, or if the device failure led to patient injury, or surgical delay.
|
|
Manufacturer Narrative
|
Udi: (b)(4).Device history record (dhr) - the dhr shows no abnormalities related to the reported failure.The reported complaint was confirmed from the evaluation of the returned mayfield skull clamp.The lock of the device was loose.Evaluation showed that the lock has rotational and lateral movement and a residue buildup is present.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|