Model Number CI-1400-01 |
Device Problems
Loose or Intermittent Connection (1371); Mechanical Problem (1384); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 12/24/2020 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing retention issues.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed a detached antenna coil.Also, the electrode was severed prior to receipt.These are believed to have occurred during revision surgery.The external visual inspection also revealed a void in the magnet weld.The device passed the photographic imaging inspection.System lock could not be performed due to the condition of the device.The device passed the electrical and mechanical's test performed.The magnet failed the gauss measurement test.The magnet was corroded due to a pin hole found in the weld which compromised magnet hermeticity.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|