MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem High Test Results (2457)
Patient Problem Hypoglycemia (1912)
Event Date 08/10/2020
Event Type  Injury  
Manufacturer Narrative
Sensor (b)(4) has been returned and investigated.Visual inspection was performed on the returned sensor, no issues were observed.Sensor plug was properly seated in the mount.Data was extracted using approved software, sensor was found to be in sensor state 6 (indicating abnormal termination).The sensor plug was removed and the plug assembly was inspected, no issues were observed.The sensor was reprogrammed, the battery was replaced, and data was extracted again, however the sensor remained in sensor state 6.An extended investigation was also performed on the returned sensor.A visual inspection on the printed circuit board assembly (pcba) revealed extensive corrosion and attempts to reprogram the sensor were unsuccessful.In order to test the complaint of high readings, a linearity test must be performed, however since the damage to the pcba was so extensive, a linearity test could not be performed, and the voltage could not be monitored.Adc was therefore unable to perform further testing related to the high readings issue reported by the customer.A separate investigation into the abnormal termination (sensor state 6) was performed (which was noted to have occurred at the time of the investigation and was not part of the customer¿s original complaint.) it was observed that the sensor was in state 6 with event code 21 (indicating a temporary drop in the source voltage) and the sensor was observed to have terminated on the body.The sensor was de-cased and corrosion was found, therefore the abnormal termination was attributed to corrosion.No further investigation is required.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A high readings issue was reported with the adc freestyle libre 2 sensor.A customer obtained readings of 475 mg/dl, 474 mg/dl, 447 mg/dl, with no comparative readings, and self-treated with fast-acting insulin (dose unknown).The customer noticed he reached low glucose levels and was taken to the hospital.The customer obtained a laboratory result of 46 mg/dl, was diagnosed with hypoglycemia, and treated with glucose drip.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
erica frank
1360 south loop road
alameda, CA 94502-7001
5104242454
MDR Report Key10901282
MDR Text Key218191605
Report Number2954323-2020-11825
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/28/2021
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Initial Date Manufacturer Received 08/11/2020
Initial Date FDA Received11/25/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age34 YR
Patient Weight93
-
-