On (b)(6) 2020, a 20mm amplatzer septal occluder was selected for implant.During the procedure, the physician unsheathed the left atrial disc and was pulling back the device in order to un-sheath the right disc.Since the device was not aligning to the septum properly, the physician decided to re-sheath the left atrial disc and rotate the sheath in effort to adjust the angle in which the device would align to have a better orientation once it was completely unsheathed.This was attempted 3 times, in which the device would not sit properly.On the 4th time, when the physician rotated the sheath to change angles, the device was deployed and formed a cobra shape.The device was retracted, removed from the patient, and exchanged for another 20mm amplatzer occluder (lot # 6674420).The second device was successfully implanted.The delivery system was not exchanged when the device was exchanged.There was no negative impact on the patient and there were no significant delays in the procedure.
|