Model Number CI-1600-04
Device Problems Mechanical Problem (1384); Impedance Problem (2950)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 12/02/2020
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing impedance issues and decreased performance.Programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed damaged silicone on the top cover, and the electrode was severed at the fantail, near the electrode ground ring, and near the array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis of the electrode pocket revealed corrosion of electrode wires.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10907370
MDR Text Key218511119
Report Number3006556115-2020-01390
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)161213(17)191231
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/31/2019
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/26/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received12/23/2020
02/08/2021
07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
-
-