It was reported a clearlink system continu-flo solution set separated, and subsequently leaked.The event was described as the tubing became disconnected at the stopcock, and, "resulted in backflow of blood" with approximately 100 ml of blood leaked from the connector of a clearlink set.There was no report of patient injury, or medical intervention associated with this event.No additional information is available.
|
The actual device was not available; however, a companion and photograph of the sample were received for evaluation.Visual inspection of the companion sample did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed on the companion sample including clear passage and pressure testing; and the device performed according to product specifications.During visual inspection of the provided photograph, the tubing was observed separated from the y-site clearlink in the downstream part.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.
|