Model Number 74121142
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Pain (1994); Joint Dislocation (2374); Metal Related Pathology (4530)
Event Date 11/07/2011
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Bilateral patient.Plaintiff underwent medically-indicated revision of the bhr right hip implants on (b)(6) 2011.The patient revision surgery was performed due to unknown reasons.Only the head was explanted.A conversion to a bhr-tha system was performed.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed on a bilateral patient.During the revision, the bhr head was removed.The bhr cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr head and bhr cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr head.Similar complaints have been identified for the bhr cup and this will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.For the bilateral two stage revisions; the clinical information provided, of the reported elevated metal ion levels, the pseudotumor, and the necrotic tissue may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure versus the reported infection.The cultures that positive for bacillus species on the left and mrse on the right were likely an hematogenous infection without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
H10: additional information added in b6 and h6 (health effect - clinical code was updated).
 
Event Description
Bilateral patient.Plaintiff underwent medically-indicated revision of the bhr right hip implants on (b)(6) 2011 due to recurrent dislocations and joint instability following a previous cup reorientation and multiple irrigation & debridement for persistent fluid accumulation in the right hip.Only the bhr resurfacing femoral head was explanted.A conversion to a bhr-tha system was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 42MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10908351
MDR Text Key218500336
Report Number3005975929-2020-00459
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502773
UDI-Public03596010502773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/26/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number74121142
Device Catalogue Number74121142
Device Lot Number086086
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-