Model Number 816300
Device Problems Calibration Problem (2890); Inadequate User Interface (2958)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2020
Event Type  malfunction  
Manufacturer Narrative
Per the perfusionist, the screen intermittently does not click where it should.A few clicks will be off by two to four inches, and then a few will be accurate.The screen was recalibrated once, but the problem was not resolved.This complaint is related to cr-79810 / mfr#: 1828100-2020-00458.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) screen calibration was intermittently off by two to three inches.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
The field service representative (fsr) verified the reported issue.He replaced the central control monitor (ccm).The unit operated to the manufacturer's specifications.During laboratory analysis, the product surveillance technician (pst) duplicated the reported compliant.It was determined that the touch screen sensor panel was defective.The suspect part was sent back to the manufacturer for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10909500
MDR Text Key218575585
Report Number1828100-2020-00463
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799001325
UDI-Public(01)00886799001325(11)200415
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/04/2020
Initial Date FDA Received11/26/2020
Supplement Dates Manufacturer Received01/21/2021
02/17/2021
Supplement Dates FDA Received02/12/2021
03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-