• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 50MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 50MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74120150
Device Problem Compatibility Problem (2960)
Patient Problems Failure of Implant (1924); Metal Related Pathology (4530)
Event Date 07/04/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a revision surgery was performed on (b)(6) 2019.The revision surgery was performed on the hip due to impending fracture femoral neck on resurfacing bhr prosthesis right side.The patient outcome is unknown.
 
Manufacturer Narrative
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information states that this event has been already reported under complaint reference (b)(4), your reference 3005975929-2019-00273, therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Manufacturer Narrative
Same incident as mdr 3005975929-2020-00472.It was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the head.A similar complaint has been identified for the cup.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.A documentation error was found in the dhr review which was investigated, this had no effect on the efficacy, quality or performance of the product.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.Smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 50MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key10913544
MDR Text Key218494151
Report Number3005975929-2020-00473
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502568
UDI-Public03596010502568
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/30/2008
Device Model Number74120150
Device Catalogue Number74120150
Device Lot Number28224
Initial Date Manufacturer Received 11/05/2020
Initial Date FDA Received11/27/2020
Supplement Dates Manufacturer Received12/09/2020
01/28/2021
Supplement Dates FDA Received12/23/2020
02/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age71 YR
-
-