Model Number N/A |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under cmp-(b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.Telephone number: (b)(6).
|
|
Event Description
|
It was reported that the dermatome stopped working during surgery.There was no harm, an unknown delay, and no additional graft was needed.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
No additional information.
|
|
Manufacturer Narrative
|
This follow up report is being submitted to report additional information.Review of the most recent repair record determined the motor was not functioning.The motor was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|