Model Number M3-30A
Device Problems Break (1069); Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation confirmed the reported issue.The meniscus at distal tip was touching the cover glass and was damaged.The light fibers were found broken.The damages observed caused debris on the undercover glass and resulted in poor image.The damages found on the device were most likely caused by user mishandling.Device history records were reviewed and showed the product met all specifications upon release.As stated on the ifu (instruction for use) and as a preventive measure, study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects olympus will continue to monitor complaints for this device.
 
Event Description
It was reported that the device was found with bent tip and sharp edges, sharp burr on the distal tip.There were no further details provided regarding the event.No patient involvement on this event.No user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Type of Device
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
93 north pleasant st.
norwalk, OH 44857
9013785969
MDR Report Key10924697
MDR Text Key222023690
Report Number1519132-2020-00101
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K890328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3-30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2020
Initial Date FDA Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-