Device evaluation confirmed the reported issue.The meniscus at distal tip was touching the cover glass and was damaged.The light fibers were found broken.The damages observed caused debris on the undercover glass and resulted in poor image.The damages found on the device were most likely caused by user mishandling.Device history records were reviewed and showed the product met all specifications upon release.As stated on the ifu (instruction for use) and as a preventive measure, study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects olympus will continue to monitor complaints for this device.
|