• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24653
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 10/27/2020
Event Type  Injury  
Event Description
It was reported that thrombosis occurred.The subject was enrolled in the (b)(6) study on (b)(6) 2017 and the index procedure was performed on the same day.The target lesion was located in the right distal superficial femoral artery (sfa) with 95% stenosis and was 140 mm long with a proximal reference vessel diameter of 6 mm and a distal reference vessel diameter of 6 mm and was classified as tasc ii b lesion.The target lesion was treated with pre-dilatation followed by a placement of a 6 mm x 150 mm study stent.Following post dilation, 5% residual stenosis was noted.On (b)(6) 2017, the subject was discharged on aspirin and clopidogrel.On (b)(6) 2020, the subject visited the enrolling site due to pain in the right calf which began approximately 4 to 5 months previously.The intermittent claudication symptoms were particularly noted while climbing.Subject additionally had ap symptoms.Physical examination revealed easily palpable arteries at dorsalis pedis bilaterally.On the same day, the subject was admitted as an inpatient for an elective pelvic/leg angiography.On (b)(6) 2020, angiography revealed stent thrombosis in the right proximal to distal sfa which was recanalized by performing percutaneous transluminal angioplasty (pta) using a non-boston scientific 5 mm x 120 mm balloon followed by non-boston scientific 5 mm x 200 mm drug-coated balloons.Additionally, coronary angiography was also performed which revealed stenosis of d1 and lateral branch with good, unimpaired left ventricular function.On (b)(6) 2020, post procedural doppler ultrasound scan revealed good interventional results with flow velocity at the right sfa at 150 cm/s.A dual antiplatelet therapy aggregation inhibition with acetylsalicylic acid (asa) and clopidogrel was recommended for 6 months.Post 6-months, asa as a mono therapy with retention of statin was suggested.A follow-up appointment was made on (b)(6) 2021.On (b)(6) 2020, the event was considered recovered/ resolved and the subject was discharged on same day in good general condition.
 
Event Description
It was reported that thrombosis occurred.The subject was enrolled in the eminent study on (b)(6) 2017 and the index procedure was performed on the same day.The target lesion was located in the right distal superficial femoral artery (sfa) with 95% stenosis and was 140 mm long with a proximal reference vessel diameter of 6 mm and a distal reference vessel diameter of 6 mm and was classified as tasc ii b lesion.The target lesion was treated with pre-dilatation followed by a placement of a 6 mm x 150 mm study stent.Following post dilation, 5% residual stenosis was noted.On (b)(6) 2017, the subject was discharged on aspirin and clopidogrel.On (b)(6) 2020, the subject visited the enrolling site due to pain in the right calf which began approximately 4 to 5 months previously.The intermittent claudication symptoms were particularly noted while climbing.Subject additionally had ap symptoms.Physical examination revealed easily palpable arteries at dorsalis pedis bilaterally.On the same day, the subject was admitted as an inpatient for an elective pelvic/leg angiography.On 16-nov-2020, angiography revealed stent thrombosis in the right proximal to distal sfa which was recanalized by performing percutaneous transluminal angioplasty (pta) using a non-boston scientific 5 mm x 120 mm balloon followed by non-boston scientific 5 mm x 200 mm drug-coated balloons.Additionally, coronary angiography was also performed which revealed stenosis of d1 and lateral branch with good, unimpaired left ventricular function.On (b)(6) 2020, post procedural doppler ultrasound scan revealed good interventional results with flow velocity at the right sfa at 150 cm/s.A dual antiplatelet therapy aggregation inhibition with acetylsalicylic acid (asa) and clopidogrel was recommended for 6 months.Post 6-months, asa as a mono therapy with retention of statin was suggested.A follow-up appointment was made on (b)(6) 2021.On (b)(6) 2020, the event was considered recovered/ resolved and the subject was discharged on same day in good general condition.It was further reported that on (b)(6) 2020, the subject visited the site for protocol scheduled 36-month follow-up.Additionally, over the past 6 months, the subject had noted right calf claudication symptoms at a walking distance of 200m, which was in the case of climbing earlier.Rutherford classification on the same day was at 2 (moderated claudication).Duplex ultrasound sonography performed on the same day revealed 2 highest progressive in-stent restenosis, indicative of re-intervention which was planned on a later date.In connection with recurring coronary stenosis, ezetrol 10mg was recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10927603
MDR Text Key218936098
Report Number2134265-2020-16696
Device Sequence Number1
Product Code NIU
Combination Product (y/n)N
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model Number24653
Device Catalogue Number24653
Device Lot Number0019834884
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/01/2020
Supplement Dates Manufacturer Received12/03/2020
Supplement Dates FDA Received12/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PHILIPS STELLAREX DRUG-COATED BALLOONS; PHILIPS STELLAREX DRUG-COATED BALLOONS; PHILIPS STELLAREX DRUG-COATED BALLOONS
Patient Outcome(s) Required Intervention;
Patient Age65 YR
-
-