ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE Back to Search Results
Model Number N/A
Device Problem Failure to Cut (2587)
Patient Problem Scar Tissue (2060)
Event Date 11/06/2020
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.(b)(6).
 
Event Description
There was a malfunction and this dermatome was not cutting properly.The patient had unforeseen scar wounds.
 
Event Description
There was a malfunction and this dermatome was not cutting properly.The event took place during surgery and an alternate device was not required to be used.The patient had unforeseen scar wounds.No additional patient consequences were reported.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d10, g4, g7, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined that the unit was out of calibration at the 0, 4, and 12 readings.The unit was recalibrated and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10930339
MDR Text Key219125801
Report Number0001526350-2020-01023
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64371029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-