• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 11/11/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report during preparation, the steerable guide catheter did not hold column.It was reported that during preparation while performing functional testing of the steerable guiding catheter (sgc), it was observed that the hemostatic valve didn't hold the water column.It seemed that the valve had a leak.A new sgc was used.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated and the reported leak was confirmed via returned device analysis and observed that the silicone fluid was missing in the valve chamber.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation determined the reported leak appears to be related to the identified missing silicone fluid in the valve chamber.The missing silicone fluid appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular (av) will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10930821
MDR Text Key219177340
Report Number2024168-2020-10067
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/25/2021
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number00623U703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2020
Initial Date Manufacturer Received 11/11/2020
Initial Date FDA Received12/01/2020
Supplement Dates Manufacturer Received02/02/2021
Supplement Dates FDA Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-