On 11/13/2020, a distributor of zyno medical reported a complaint.A user facility representative contacted the distributor on 11/12/2020, and stated that an administration set model#: b2-70072; lot#: 20055819, had a pinch clamp that was upside down.No patient was involved.No medication was being infused.The device operator was a registered nurse.On 11/23/3030, the distributor confirmed that the end-user only found one administration set with the inverted pinch clamp in the entire case and that the affected administration set will not be returned.The distributor also provided a picture of the inverted pinch clamp.On 11/24/2020, the distributor provided two additional pictures of the affected set.The contract manufacturer of the affected device is becton, dickinson and company.
|