(b)(6).The actual device was not available; however, six (6) retained samples were evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Two of the samples underwent underwater leak testing and no leaks were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|