(b)(6).Device evaluated by mfr.: a promus premier ous mr 16 x 3.00mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found the tip detached and not returned with the device.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.The device was loaded onto a 0.0140" guidewire and then attached to encore inflation device.The balloon was inflated to 16 atm and the stent was successfully expanded.A vacuum was pulled using the encore device and the balloon delated fully.The encore device was verified before and after the procedure.No other issues were identified during the product analysis.
|