BOSTON SCIENTIFIC CORPORATION INTELLANAV STABLEPOINT OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Lot Number 0025958733 |
Device Problems
Communication or Transmission Problem (2896); Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported that a intellanav stable point open-irrigated catheter was used in a right typical flutter procedure.When the catheter was introduced in patient's body, maestro was not able to detect the catheter.The maestro pod, rf generator and rhythmia were restarted without solving the issues.After several attempts, the catheter was finally detected by the rf generator but temperature measured on the maestro was 67 degrees.No errors were displayed on the maestro console despite the high temperature.The procedure was completed by replacing the catheter with no patient complications being reported.
|
|