Model Number 176625 |
Device Problems
Mechanical Problem (1384); Failure to Align (2522); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, pre-operatively, there were any issues experienced with the loading or firing of the clips.There was no patient involved.
|
|
Manufacturer Narrative
|
This report 9612501-2020-01788 was sent in error as a duplicate for mfr report number: 9612501-2020-01590, any additional information received in the future will be captured and submitted under the report number 9612501-2020-01590.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
According to the reporter, pre-operatively, there were any issues experienced with the loading or firing of the clips.The surgeon figured out that the jaws out of shape so the device was tested and found out that the clips could not form properly.There was no patient involved.
|
|
Manufacturer Narrative
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the handle was broken and out of position in a manner consistent of cycling through the safety interlock.It was reported that the clips did not load properly into the jaws as expected, did not close completely and were malformed.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if an attempt was made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature was designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt was made to forcibly fire the instrument while engaged in interlock, the lugs were designed to break as noted and the interlock feature continues to function as intended.It was also reported that the jaws were misaligned.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|