• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Model Number 500101-001
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
The cpc connector has been returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the spring in the cpc connector was displaced.The cpc connector was exchanged without any difficulty.There was no reported adverse patient impact.
 
Manufacturer Narrative
Visual inspection confirmed the customer-reported issue of a displaced spring, as the spring inside the left (female) cpc connector was displaced inside the housing beneath the metal release tab.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its investigation and is closing this file.Ce 5415 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10945164
MDR Text Key219514447
Report Number3003761017-2020-00240
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number119214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2020
Date Manufacturer Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age32 YR
-
-