On (b)(6) 2020, a 16 mm amplatzer septal occluder was chosen for a procedure.During the procedure, the user reported the left atrial side of the device did not correctly deploy from the sheath.The physician tried to recapture and redeployed the device approximately four times, but the left disc remained round each time, in a bulbous shape.The physician exchanged the device and completed the procedure with no further events.
|
The reported event of the device deploying deformed could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.Please note, per the amplatzer septal occluder instructions for use, artmt100116885 revision a "caution: do not release the device from the delivery cable if the device does not conform to its original configuration or if device position is unstable or interferes with any adjacent cardiac structure (such as svc, pv, mv, cs, ao).Advance the delivery sheath to recapture the device and redeploy.If the device position is still unsatisfactory, recapture the device and replace it with a new device or abort the procedure.".
|