Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Injury (2348)
Event Date 11/12/2020
Event Type  Injury  
Manufacturer Narrative
Unique device identifier (udi): (b)(4).The perforator was not returned for evaluation (per customer, product not available) therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
A physician reported that on (b)(6) 2020 the perforator was used on a (b)(6)-year-old male patient with moyamoya disease.The perforator failed to disengage causing a dura mater injury that was treated with duragen.No information regarding surgical delay was provided.No further information was provided by the hospital.
 
Manufacturer Narrative
Updated fields:  d4, d9, g1, g3, g6, h2, h3, h4, h6, h10.The perforator was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.Failure analysis - the perforator unit was inspected using the unaided eye.Unit had damaged/illegible "eo" label, no additional anomalies.The ifu testing procedure was performed with no observed anomalies.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.In the failure analysis that was performed, the returned unit was found to work as intended, and met all acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.A dhr review and trending were performed as part of the evaluation.Proper finished goods testing was performed prior to release as indicated in the dhr.Product was received for analysis and the investigation could not confirm the complaint.The risk remains acceptable per the risk analysis.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10949940
MDR Text Key220687732
Report Number3014334038-2020-00115
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ7045Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2021
Initial Date Manufacturer Received 11/12/2020
Initial Date FDA Received12/04/2020
Supplement Dates Manufacturer Received03/22/2021
Supplement Dates FDA Received04/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRIMADO DRILL-ROTATION NUMBER 80000