Catalog Number A2114 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the mayfield infinity skull clamp (product id a2114) was broken.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
|
|
Manufacturer Narrative
|
Udi: (b)(4).The reported complaint was confirmed from the evaluation of the returned mayfield infinity skull clamp.Unit received without pedi rocker arm or suit case.Evaluation showed that the left and right pawls of the device are worn.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.This device exceeds its expected life of 7 years (manufactured in 2012).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|