As part of our investigation, on november 16, 2020 an olympus field service engineer (fse) was dispatched to the customer¿s site to service the aer.The fse reported that the lid was cracked in the left rear corner and was replaced.The lid was removed per the oer-pro technical guide and confirmed the procedure by using the attached checklist.The fse replaced the alcohol connector that was broken when the alcohol was added.Also, the oer¿s drain hose was replaced as it was leaking due to the customer moving the oer-pro to clean.The software attributes were verified and confirmed.The covers of the oer-pro were not removed so an electrical safety check was not required.The fse reported that the cause of the cracked lid is unknown.The oer-pro instruction manual provides the following statements to prevent damage to the oer-pro.Before using the equipment, always check that there is no irregularity regarding the following points on the lid and the lid packing.If there is any irregularity, cleaning fluid or disinfectant solution may leak out.The lid is not cracked, broken, or otherwise damaged.When closing the lid, be careful not to get the connecting tube caught between the reprocessing basin and lid and make sure the endoscopes and the washing case are not touching the lid.Close the lid after ensuring that the cover of the washing case is closed.If the endoscopes and the washing case are touching the lid, adjust their positions and close the lid.If the lid is closed while contacting the endoscopes or accessories such as the washing case and the connecting tubes, the endoscopes, the accessories and the equipment may get damaged or water leakage may result.
|