Model Number N/A
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
The quality control (qc) results were within the acceptable range and the calibration was acceptable.Siemens healthcare diagnostics has concluded its investigation of atellica im and advia centaur sars-cov-2 total (cov2t) non-reproducible false reactive results with kit lots ending in 004, 005, 035 and 006.Siemens investigation determined that although non-reproducible false reactive results were observed with multiple kit lots, results indicate the negative percent agreement confidence interval of the kit lots evaluated overlap the confidence interval in the ifu; therefore they are performing within claims and a change in performance has not been confirmed.The ifu states in the limitations section: "this assay should not be used to diagnose or exclude acute infection.Results are not intended to be used as the sole basis for patient management decisions.Test results should be interpreted in conjunction with clinical observations, patient history, epidemiological information, and other laboratory findings.A reactive test result does not exclude past or present infection by other coronaviruses, such as sars cov-1, mers-cov, hku1, 229e, nl63, or oc43." a false positive/reactive result would not be used in isolation and would be correlated with clinical history.Additional laboratory testing would be used to determine specific antibody presence.Although there is no potential for serious injury in this case, an mdr will be reported to the fda as a requirement of the emergency use authorization (eua).The advia centaur sars-cov-2 total (cov2t) lot 035 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.
 
Event Description
The customer obtained a reactive (positive) advia centaur xpt sars-cov-2 total (cov2t) result for one patient.The initial reactive (positive) result was considered discordant when compared to the nonreactive (negative) repeat results.It is unknown which result the customer reported to the physician.There are no known reports of patient intervention or adverse health consequences due to the discordant cov2t result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XPT SARS-COV-2 TOTAL (COV2T)
Type of Device
SARS-COV-2 IMMUNOASSAY,
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC
333 coney street
east walpole MA 02032
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
5086604381
MDR Report Key10971990
MDR Text Key247454721
Report Number1219913-2020-00617
Device Sequence Number1
Product Code QKO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EUA201370
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2021
Device Model NumberN/A
Device Catalogue Number11206710
Device Lot Number035
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age40 YR
-
-