The quality control (qc) results were within the acceptable range and the calibration was acceptable.Siemens healthcare diagnostics has concluded its investigation of atellica im and advia centaur sars-cov-2 total (cov2t) non-reproducible false reactive results with kit lots ending in 004, 005, 035 and 006.Siemens investigation determined that although non-reproducible false reactive results were observed with multiple kit lots, results indicate the negative percent agreement confidence interval of the kit lots evaluated overlap the confidence interval in the ifu; therefore they are performing within claims and a change in performance has not been confirmed.The ifu states in the limitations section: "this assay should not be used to diagnose or exclude acute infection.Results are not intended to be used as the sole basis for patient management decisions.Test results should be interpreted in conjunction with clinical observations, patient history, epidemiological information, and other laboratory findings.A reactive test result does not exclude past or present infection by other coronaviruses, such as sars cov-1, mers-cov, hku1, 229e, nl63, or oc43." a false positive/reactive result would not be used in isolation and would be correlated with clinical history.Additional laboratory testing would be used to determine specific antibody presence.Although there is no potential for serious injury in this case, an mdr will be reported to the fda as a requirement of the emergency use authorization (eua).The advia centaur sars-cov-2 total (cov2t) lot 035 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.
|