Model Number AIA-2000
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2020
Event Type  malfunction  
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing printouts and was unable to reproduce error.Fse resolved the complaint by cleaning off the hybrid arm rail and performed a version up.Fse validated instrument by successfully completing quality control run without error and results were within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The aia-2000 operator's manual under appendix 4: error messages states the following: [4241] hybrid arm y-axis home detection failure cause: the home sensor failed to activate after the hybrid arm y-axis moved toward the home position.If retry fails, the measurement result will be flagged (mf flag).Solution: contact tosoh service center or local representatives.The most probable cause of the reported event was due to dirty hybrid arm rail.
 
Event Description
A customer reported getting error message "4241 hybrid arm y-axis home detection failure" on the aia-2000 instrument.The customer performed an all set home function and rebooted instrument, but error persisted.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
AIA-2000
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586-23
JA   1058623
MDR Report Key10972331
MDR Text Key220972109
Report Number8031673-2020-00362
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000
Device Catalogue Number022101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-