The field service engineer (fse) conducted a site visit and was able to confirm the reported error via the error log.The fse was able to reproduce the reported issue by performing a prime wash operation.The fse found the b/f probe sensitivity adjustment to be out of adjustment.The fse adjusted the b/f probe sensitivity for proper sensing.The aia-900 analyzer is functioning as expected.No further action required by field service.A 13-month complaint and service history review for serial number (b)(4) from 09oct2019 through aware date (b)(6) 2020 was performed for similar complaints.There were two similar complaints including this one identified during the search period.The aia-900 operator's manual under section 12 - flags and error messages states the following: [2236] bf probe 2 suction failure.Cause: the overflow sensor 2 s133 detected liquid after a washer suction operation a wu flag will be attached to the measurement result.Action: clean the wash probe 2.Check s133, the waste liquid solenoid valve sv171, the waste liquid tube, and the liquid pump lp173.The most probable cause of the reported event was that the b/f probe sensitivity needed adjustment.
|
Customer called to report error 2236, bf probe 2 suction failure on the aia-900 analyzer.Customer states this is an ongoing issue and that the field service engineer (fse) has been on site.Customer was told that if it happened again, to call for service.A field service engineer (fse) was dispatched to address the reported issue which caused a delay in reporting luteinizing hormone (lh ii) and follicle stimulating hormone (fsh) patient samples.There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|