• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Mechanical Problem (1384); Impedance Problem (2950)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 12/17/2020
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing impedance issues and decreased performance.Programming adjustments were made, however, the issue did not resolve.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts to the top and bottom covers and electrode.Able to see cut wires in electrode between fantail and electrode ground ring.Electrode ground ring has been displaced.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut wires in electrode between fantail and electrode ground ring.The photographic imaging inspection revealed an electrode was broken within the electrode pocket.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The scanning electron microscopy analysis of the electrode pocket revealed corrosion of electrode wires.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10976951
MDR Text Key221249084
Report Number3006556115-2020-01456
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)161201(17)191130
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2019
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/09/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received01/29/2021
06/13/2021
07/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age72 YR
-
-