It was reported that on (b)(6) 2020, ventricular tachycardia was detected by the implantable cardioverter defibrillator device but the program setting on the device was to monitor, so no therapy was delivered.Also, on (b)(6) 2020, pacing was confirmed, but the atria and ventricle were not captured.On (b)(6) 2020, patient presented in clinic for a device check.Upon review, the implantable cardioverter defibrillator exhibited noise.The reason for the noise was unclear.On (b)(6) 2020, the patient deceased.The physician commented the cause of death seemed to be an irregular pulse.The programming setting on the device was 1 zone setting, the reason was unknown, and the therapy was not delivered due to no setting zone.The patient might have passed away due to continued ventricular tachycardia.
|