SMITH & NEPHEW, INC. GII DCF VARIABLE SZ GUIDE LT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440007 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Case (b)(4).
|
|
Event Description
|
It was reported that a genesis ii dcf variable size guide left was loosened and was not able to keep correct external rotation.This occurred during a tka procedure, instrument inside the patient.No harm or injury reported on patient.Procedure was finished with the same device.No delay reported.
|
|
Manufacturer Narrative
|
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The dial of the device is heavily damaged rendering it inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|