Catalog Number VENUM14120
Device Problems Material Deformation (2976); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.
 
Event Description
It was reported that during a stent placement, the device allegedly deployed partially and difficult to remove.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: d4 (medical device catalog).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Event Description
It was reported that during a stent placement through common and external iliac vein, the device allegedly failed to expand in the distal end and stent elongated.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the lot number was not reported, a review was not performed.Previous investigations of the similar complaints have been reviewed, and similar complaints were found.Investigation summary: the physical sample was not available, and images were not provided.The alleged issue could not be re produced which led to an inconclusive evaluation result.Based on the information available the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The ifu was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H11: h6 (method, result).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a stent placement through common and external iliac vein, the device allegedly failed to expand in the distal end and stent elongated.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Type of Device
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key10982179
MDR Text Key220695234
Report Number9681442-2020-20070
Device Sequence Number1
Product Code QAN
Combination Product (y/n)N
PMA/PMN Number
P180037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVENUM14120
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/13/2020
Initial Date FDA Received12/10/2020
Supplement Dates Manufacturer Received12/11/2020
04/06/2021
Supplement Dates FDA Received01/07/2021
04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age40 YR
-
-