Model Number 60-05-65
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/12/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 erc tubing clamp / 620mm.The incident occurred in (b)(6).The affected device was requested back to the manufacturer site for investigation.Deviation could be reproduced during the investigation and an error code was displayed.Root cause has been identified as residues of dried fluids inside the clamp leading to internal parts corrosion.Use condition (improper cleaning procedure by the user) may have contributed to fluid ingress into the device.
 
Event Description
Livanova (b)(4) received a report that a s5 erc tubing clamp / 620mm was giving an error message after a procedure.The user tried to reseat the cables and restart the device.However, alarm persisted.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ERC TUBING CLAMP / 620MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada 80004
MDR Report Key10982443
MDR Text Key220642174
Report Number9611109-2020-00675
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-65
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/11/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-