Model Number 3820 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/25/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was fractured.A 06/2.75 flextome cutting balloon was selected for use.Upon unpacking, it was noted that the device was fractured in the middle.The procedure was completed with another of the same device.No complications reported and patient was stable post procedure.
|
|
Event Description
|
It was reported that the device was fractured.A 06/2.75 flextome cutting balloon was selected for use.Upon unpacking, it was noted that the device was fractured in the middle.The procedure was completed with another of the same device.No complications reported and patient was stable post procedure.It was further reported that the balloon and shaft extrusion portion of the device that was not returned, were confirmed to be discarded in the facility.The device was completely removed from the patient's body.
|
|
Event Description
|
It was reported that the device was fractured.A 06/2.75 flextome cutting balloon was selected for use.Upon unpacking, it was noted that the device was fractured in the middle.The procedure was completed with another of the same device.No complications reported and patient was stable post procedure.It was further reported that the balloon and shaft extrusion portion of the device that was not returned, were confirmed to be discarded in the facility.The device was completely removed from the patient's body.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was tightly folded.No damage or any issues were noted with the balloon material that could have contributed to the complaint incident.The markerbands and blades and tip section of the device were visually and microscopically examined, and no issues were noted with the markerbands, blades or tip of the device.All blades were present and fully bonded to the balloon material.A visual and tactile examination found that the device was returned without the shaft polymer extrusion and balloon portion of the device.72.3cm of device was returned (from strain relief to break point).Multiple kinks were also noted along several locations of the hypotube shaft.This type of damage is consistent with excessive force that could have been applied to the delivery system.The shaft polymer extrusion section of the device was not returned for analysis.
|
|
Search Alerts/Recalls
|