BALLARD* CLOSED SUCTION SYSTEM FOR NEONATES/PEDIATRICS, 6 F, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES
|
Back to Search Results |
|
Model Number 196-4J |
Device Problem
Suction Failure (4039)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is reported to be available, but has not yet been received by the manufacturer.The device history record for lot 30047544 was reviewed and the product was produced according to product specifications.All information reasonably known as of (b)(6) 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
|
Event Description
|
It was reported that the irrigation port did not suction saline.This was discovered while not suctioning the patient.The suction catheter was replaced for new one.No patient injury.No further information provided.
|
|
Manufacturer Narrative
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 04 jan 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
One used closed suction catheter was returned for evaluation.Two detached irrigation ports were returned, though there was no indication which one was original to the device.One empty amsina 5ml saline vial was also returned.The suction catheter was tested with both irrigation ports and the amsina saline vial.The device worked as expected.The lab was unable to recreate the malfunction reported by the facility.The complaint could not be confirmed as reported.No root cause could be determined.All information reasonably known as of (b)(6) 2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|
|