Catalog Number CX-FX15RW40V
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
Udi not required for product code.Implanted date: device was not implanted; explanted date: device was not explanted; occupation: clinical engineer; pma/510(k)- k130280.The actual sample was received for evaluation.Visual inspection revealed that the lock adapter had come off the male connector of the sampling system.Magnifying inspection of the actual sample did not find any visible anomaly, including deformity, in the male connector, or in the lock adapter.The outer diameter of the rib of the male connector and the inner diameter of the lock adapter were measured.Compared to a current product sample, no difference was observed in the measured values.The surface of the male connector was subjected to elemental analysis by sem-edx (scanning electron microscope / energy dispersive x-ray spectroscopy).The result showed the presence of si, which is likely to be derived from the silicone applied to the switch cocks of the sampling system to improve the lubricity of them in the manufacturing process.Reproductive testing was performed, and silicone was applied to a male connector of a factory-retained sampling system, a female connector was connected to it, and then a lock adapter was tightened up.As a result, the lock adapter came off.The factory-retained female connector was connected to the male connector of the actual sample, and then the sampling system was filled with normal saline and pressurized with syringe at 1000 mmhg.As a result, no leak occurred.Subsequently, while the lock adapter of the actual sample was put in the state of having got over the rib of the male connector, the sampling line was filled with normal saline.As a result, leak was observed at the joint between the male and female connector.From this, it was presumed that the joint between the male and female connectors became loose due to the dislodgement of the lock adapter, which resulted in the leak.A review of the device history record and product-release judgement record of the involved product code/lot# combination was conducted with no findings.Ifu states; do not use if the package or device is damaged (e.G.Cracked) or any of the port caps are off.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It is likely that the silicon, which is applied to the switch cocks of the sampling system to improve the lubricity, was transferred to the male connector part due to some factors.Afterward, the lock adapter, when re-tightened, may have got over the rib of the male connector in lubricated state and came off.However, from the available information including the state of the actual sample, it could not be determined when silicone was transferred to the male connector.The exact cause of the reported event cannot be definitively determined based on the available information.(b)(4).
 
Event Description
The user facility reported that the involved capiox device was used pre-treatment.During priming, leak occurred at the luer locking part of the sampling system and the lock adapter came off.The sampling system was changed out to cope with the problem.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX FX OXYGENATOR
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
theresa mussaw
reg. no. 2243441
265 davidson ave suite 320
somerset, NJ 08873
8002837866
MDR Report Key11001745
MDR Text Key221270825
Report Number9681834-2020-00251
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Catalogue NumberCX-FX15RW40V
Device Lot Number200429
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-