ST. JUDE MEDICAL 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY
|
Back to Search Results |
|
Model Number RFG-NT-2000 |
Device Problems
Thermal Decomposition of Device (1071); Device Emits Odor (1425); Noise, Audible (3273)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results, method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During the procedure, when switching on the system, a high pitch noise and burning smell occurred.There was no visible smoke or signs of physical fire.At the time of the event the patient had the grounding pads on and the electrodes were attached to the system.The system was power cycled and when powered back on the noise was gone but a red error code appeared on the unit.The procedure was aborted due to these issues.
|
|
Manufacturer Narrative
|
Additional information: concomitant medical products, pma/510k, if follow-up, what type, device evaluated by mfr.One nt 2000 ix neurotherm rf generator,, was received for evaluation.The nt generator was connected to an external power source and the generator powered on successfully.When trying to connect the electrode at all channel two the generator was emitting an alarm beep and could not function.The field complaint could be confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No rework or non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the cause for the reported event was isolated to a thermal event located on a capacitor of the rf control board.
|
|
Search Alerts/Recalls
|
|
|