• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-2000
Device Problems Thermal Decomposition of Device (1071); Device Emits Odor (1425); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2020
Event Type  Injury  
Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the procedure, when switching on the system, a high pitch noise and burning smell occurred.There was no visible smoke or signs of physical fire.At the time of the event the patient had the grounding pads on and the electrodes were attached to the system.The system was power cycled and when powered back on the noise was gone but a red error code appeared on the unit.The procedure was aborted due to these issues.
 
Manufacturer Narrative
Additional information: concomitant medical products, pma/510k, if follow-up, what type, device evaluated by mfr.One nt 2000 ix neurotherm rf generator,, was received for evaluation.The nt generator was connected to an external power source and the generator powered on successfully.When trying to connect the electrode at all channel two the generator was emitting an alarm beep and could not function.The field complaint could be confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No rework or non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the cause for the reported event was isolated to a thermal event located on a capacitor of the rf control board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR
Type of Device
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
MDR Report Key11002526
MDR Text Key221246028
Report Number2184149-2020-00204
Device Sequence Number1
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (y/n)N
PMA/PMN Number
K111576
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number6784193
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2184149-04/22/21-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-