TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801763 |
Device Problems
Break (1069); Electrical /Electronic Property Problem (1198)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/19/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heart lung machine (hlm) had a damaged power cord.One of the plug prongs was missing.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Updated blocks: d9, h3 and h6.The reported complaint was confirmed.The field service representative (fsr) verified the reported complaint as the plug end was already replaced by the user facility's biomedical technician.The originally damaged plug end was discarded.The fsr replaced the entire power cord, including the plug end.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|