• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to replace the damaged lid and quick connect on three port valves.The device passed all the service inspections.Software attributes have been verified and confirmed.No other issues were reported.If additional information is obtained a supplemental report will be filed.
 
Event Description
A user facility reported that during preventive maintenance on an endoscope reprocessor, a cracked lid was found.There was no leaks associated with the cracked lid.No patient injuries or infections were reported.No additional information has been obtained.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record confirmed the subject device was shipped in accordance to specifications.Based on the results of the investigation, the phenomenon likely occurred due to the lid being in contact with a scope, something hard hitting the lid, or chemical crack due to adhered chemical solution which was not removed.A review of the the instructions for use that confirms the phenomenon remains detectable: 3.2: before using the equipment, always check that there is no irregularity regarding the following points on the lid and the lid packing.If there is any irregularity, cleaning fluid or disinfectant solution may leak out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11005932
MDR Text Key227200026
Report Number8010047-2020-10328
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/19/2020
Initial Date FDA Received12/14/2020
Supplement Dates Manufacturer Received01/19/2021
Supplement Dates FDA Received02/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-