Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that after a successful cryo ablation procedure, the sheath was removed from the patient and a kink and broken shaft was observed.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath 4fc12 with lot number 0010396224 was returned and analyzed.Visual inspection showed the shaft was kinked and twisted at 1.86 inches from the tip.The shaft tip was intact with no apparent issues.The kink could cause deflection or steerability or insertion difficulties.In conclusion, the reported kink issue was confirmed through testing.The sheath failed the returned product inspection due to the kinked and twisted shaft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key11006174
MDR Text Key221542646
Report Number3002648230-2020-00664
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2022
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010396224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2020
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/15/2020
Supplement Dates Manufacturer Received12/24/2020
Supplement Dates FDA Received01/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-