• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2020
Event Type  Injury  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The additional device referenced is filed under a separate medwatch report number.
 
Event Description
This is filed to report leak and medical intervention it was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3+.The first steerable guide catheter (sgc 01006u247) was advanced to the mitral valve; however, the dilator lost fluid column after dilator removal.Therefore, additional aspirations were performed and the sgc was removed.A second sgc (00822u211) was advanced to the mitral valve; however, the second sgc also lost fluid column after dilator removal.Additional aspirations were performed.A new sgc was not available, and therefore a decision was made to exchange the dilator for the second sgc.The sgc was successfully used and air never entered the patient.One clip was implanted, reducing mr to less than 1.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated and the loss of fluid column (leak) could not be confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the loss of fluid column (leak).The additional therapy/non-surgical treatment was a result of case specific circumstances as additional aspiration was performed.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11008206
MDR Text Key221503530
Report Number2024168-2020-10469
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/21/2021
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number00822U211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Initial Date Manufacturer Received 11/19/2020
Initial Date FDA Received12/15/2020
Supplement Dates Manufacturer Received01/14/2021
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age89 YR
Patient Weight49
-
-