The long term loaner device was previously returned to pentax medical from a customer on 19-nov-2020 and inspection of the unit was performed under service order (b)(4) where the quality control inspector documented the following codes on 20-nov-2020: distal cap - fixed type failed epoxy seal integrity inspection, air/ water nozzle clogged with inorganic debris, up/ down control knob/ lever markings faded, istal tip annual maintenance, passed dry leak test, passed wet leak test, elevator body sticking, air/ water socket cylinder o-ring chipped, prism scratched, right/ left control knob markings faded, fluid invasion not observed in pve connector, bending rubber mild discoloration, light carrying bundle distal cover glass single chipped, sluggish air delivery function, sluggish water delivery function, fluid invasion not observed in control body, image mild spot.The duodenoscope's repairs will include the distal case/cap, which will be replaced and/or resealed pursuant to the field correction, along with the following parts, and returned to the user upon completion: o-rings and seals, deflector operating wire, lcb distal cover, objective prism assy, distal case/cap, o-ring(0.5x1.3) imp-1, segment staycoil assy imp-c/pb-free, bending rubber.The endoscope is awaiting repair or approval by final qc as of 14-dec-2020.
|