SIEMENS HEALTHCARE DIAGNOSTICS, INC. ATELLICA IM SARS-COV-2 TOTAL (COV2T); SARS-COV-2 IMMUNOASSAY
|
Back to Search Results |
|
Model Number N/A |
Device Problem
False Positive Result (1227)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The ifu states in the limitations section: "this assay should not be used to diagnose or exclude acute infection.Results are not intended to be used as the sole basis for patient management decisions.Test results should be interpreted in conjunction with clinical observations, patient history, epidemiological information, and other laboratory findings.A reactive test result does not exclude past or present infection by other coronaviruses, such as sars-cov-1, mers-cov, hku1, 229e, nl63, or oc43." a false positive/reactive result would not be used in isolation and would be correlated with clinical history.Additional laboratory testing would be used to determine specific antibody presence.Although there is no potential for serious injury in this case, an mdr will be reported to the fda as a requirement of the emergency use authorization (eua).Siemens healthcare diagnostics is investigating.
|
|
Event Description
|
The customer performed a correlation study comparing the initial reactive results to the repeat results on the atellica im s/n (b)(4) and atellica im s/n (b)(4) with reagent lot 005.The customer observed discordant results.The initial results were reactive, and the repeats were nonreactive or reactive and nonreactive.Corrected reports were reported to the physicians.There are no known reports of patient intervention or adverse health consequences due to the discordant cov2t results.
|
|
Manufacturer Narrative
|
Siemens filed the initial mdr 1219913-2020-00630 on december 14, 2020.December 29, 2020 additional information: siemens healthcare diagnostics has concluded its investigation of atellica im and advia centaur sars-cov-2 total (cov2t) non-reproducible false reactive results with kit lots ending in 004, 005, 035 and 006.Siemens investigation determined that although non-reproducible false reactive results were observed with multiple kit lots, results indicate the negative percent agreement confidence interval of the kit lots evaluated overlap the confidence interval in the ifu; therefore they are performing within claims and a change in performance has not been confirmed.The ifu states in the limitations section: "this assay should not be used to diagnose or exclude acute infection.Results are not intended to be used as the sole basis for patient management decisions.Test results should be interpreted in conjunction with clinical observations, patient history, epidemiological information, and other laboratory findings.A reactive test result does not exclude past or present infection by other coronaviruses, such as sars cov-1, mers-cov, hku1, 229e, nl63, or oc43." a false positive/reactive result would not be used in isolation and would be correlated with clinical history.Additional laboratory testing would be used to determine specific antibody presence.The atellica im sars-cov-2 total (cov2t) lot 005 is performing as intended.A product performance issue has not been identified.No further evaluation of the device is required.Mdrs 1219913-2020-00629 supplemental report 1 through 1219913-2020-00631 supplemental report 1 were filed for different dates of testing.
|
|
Search Alerts/Recalls
|
|
|