Model Number TJF-260V
Device Problem Mechanical Problem (1384)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that in unspecified timing the forceps elevator remained been raised.Olympus service operation repair center (sorc) checked the subject device and found that the reported phenomenon could not duplicated.There was no patient injury associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacture history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus service operation repair center (sorc), there was the possibility that the reported phenomenon was attributed to a foreign matter got stuck in the forceps elevator, an external factor, or user handling.
 
Manufacturer Narrative
This supplemental report is being submitted to provide correction of ¿d9¿, ¿h3¿, ¿h6¿ and ¿h10¿ in the initial report and the follow-up number 1 report (the correct answer is please see each item of this report).The subject device was not returned to olympus medical systems corp.(omsc) for evaluation but was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon could not be duplicated.Based upon the information from sorc, omsc considered that the reported phenomenon was attributed to temporary pinching of foreign materials due to external factors or handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11010706
MDR Text Key221558918
Report Number8010047-2020-10379
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberTJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received12/15/2020
Supplement Dates Manufacturer Received01/07/2021
05/10/2021
Supplement Dates FDA Received01/14/2021
09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-