Model Number TJF-260V |
Device Problem
Mechanical Problem (1384)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not been returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed from the user that in unspecified timing the forceps elevator remained been raised.Olympus service operation repair center (sorc) checked the subject device and found that the reported phenomenon could not duplicated.There was no patient injury associated with this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacture history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus service operation repair center (sorc), there was the possibility that the reported phenomenon was attributed to a foreign matter got stuck in the forceps elevator, an external factor, or user handling.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide correction of ¿d9¿, ¿h3¿, ¿h6¿ and ¿h10¿ in the initial report and the follow-up number 1 report (the correct answer is please see each item of this report).The subject device was not returned to olympus medical systems corp.(omsc) for evaluation but was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon could not be duplicated.Based upon the information from sorc, omsc considered that the reported phenomenon was attributed to temporary pinching of foreign materials due to external factors or handling.
|
|